34 — The trend from the readings: which math

2026-11-14 · investigation report · issue #204 · manufacturers’ guides and patents (Dexcom, Abbott, Medtronic, Sanvita), literature (Facchinetti / Sparacino / Cobelli 2010, Aleppo 2017, Kudva 2018, Eichenlaub 2024, Kovatchev 2005, Dunn 2004, 21 CFR 862.1355), the sources of xDrip+, xDrip4iOS, Nightscout, oref0, AndroidAPS, Loop, Trio, Juggluco and GlucoDataHandler, user discussions · nine candidates prototyped and evaluated on the owner’s own 1-min readings, locally and read-only (investigations/204-trend); this report holds aggregate numbers only

The arrow beside the value answers “where is it going?”, the question behind most decisions about food and insulin. Today it comes from the sensor’s own rate (Abbott’s, for the Libre 3), which lags: on 2026-11-14 the readings climbed about 1 mg/dL a minute while the sensor said 0.79 and the arrow showed →; on the #193 evening it still said “falling” minutes into the rebound. The owner decided the trend is worked out from the readings by default (the sensor’s rate stays a setting and the fallback), and asked: simply subtracting the glucose from x minutes ago seems like a bad method; what math should it be? This investigation compares the methods used by manufacturers, research and DIY apps on the owner’s real readings and recommends one, with its parameters, the text bands’ hysteresis, the drawn arrow’s angle and the shared test vectors.

  • Recommendation: a least-squares slope weighted toward the newest readings: weights exp(−age / τ) over the last 5τ, τ = 3 min (15 min of readings), at least two readings spanning at least 3 min, starting over after a step longer than 12.5 min or a LO / HI. It is a pure function of the readings (no state: a restart, a backfilled reading or a follower with the same readings gets the same arrow), reads a straight line exactly, needs no resampling and uses the same numbers for 1-min and 5-min sensors. About 15 multiply-adds per reading.
  • On the owner’s 4.6 days of 1-min readings it reaches ±1 mg/dL/min 4 min after the hindsight slope (p90 5 min); Abbott’s rate takes 7 min (p90 17 min) and misses half the crossings: its rate is only about a third of how fast the readings move (gain 0.35). The two-point delta (Nightscout, xDrip+) is 1–2 min quicker but changes band twice as often and falls apart with noise; Loop’s 15-min regression is 3 min slower; the Savitzky–Golay end point is the noisiest; Holt, a Kalman filter and AndroidAPS’ shortAvgDelta score the same as the recommendation (they are nearly the same filter) but are stateful or need a 5-min grid.
  • Presets, if wanted: Quick τ 2 / Balanced τ 3 (default) / Smooth τ 5, with “From the sensor” as the fourth choice of the same setting. Text bands with hysteresis 0.2 mg/dL/min (−20 % band changes, +0.7 min).
  • The drawn arrow: linear, 0° at 0, vertical at ±3 mg/dL/min (the PoC), no extra smoothing, a short spring between readings. Doubled from 3, with the same hysteresis.
  • Two consequences for the owner (§7): the readings move ≥ 2 mg/dL/min far more often than Abbott’s rate says (~25 times a day instead of ~2): ↑ / ↓ and ⇈ / ⇊ will be seen much more, and the rate alerts can’t simply switch to this trend. The #193 quiets keep their own rule: nothing here beats it for what it does.

Manufacturers (the algorithms are not published):

Who What is known Source
Dexcom bands ±1 / 2 / 3 mg/dL/min (G7: “< 30 / 30–60 / 60–90 / > 90 mg/dL in 30 minutes”); “trend arrows are determined by recent rtCGM measurements (generally the most recent 10 minutes of glucose values)”; the G7 transmitter sends the rate itself (signed byte ÷ 10 mg/dL/min) guides G7, G6 (wording from search excerpts, not checked against the PDFs); Aleppo 2017; G7SensorKit G7GlucoseMessage.swift L33-52
Dexcom patent “maximal sustained rate of change … about 4 to 6 mg/dl/min”, “maximum sustained acceleration … about 0.1 to 0.2 mg/min/min”, used to reject outliers, not for the arrow US8954128B2
Abbott Libre five arrows at ±1 / 2; “The trend arrows are calculated from glucose readings with an emphasis on the most recent 15 minutes”; the Libre 3 sends rateOfChange (int16 ÷ 100) Kudva 2018; Juggluco libre3/bluetooth.cpp L134-151; how it smooths: not found
Medtronic patent “SG smoothed rate of change … using the 5 most-recent Isig values”, weights [0.2, 0.1, 0, −0.1, −0.2] (a 5-point least-squares slope) US11253176 (excerpt)
Sanvita patent rate “based on five consecutive data points”, bands 1 / 2 / 3 US11197627B2

Research and regulation:

  • Kalman / Bayesian smoothing: Facchinetti, Sparacino, Cobelli, An online self-tunable method to denoise CGM sensor data, IEEE TBME 2010;57(3):634 (and TBME 2011;58(9):2664). Their review: “the more the filtering, the larger the delay”; “the delay introduced by KF is about 35% less than that obtained by MA”; hypo alerts by “first-order linear extrapolation of glucose obtained from the last two/three samples” (Sparacino 2010, PMC3231130). Their filter is tuned per person and per sensor noise; it smooths the value, the rate falls out of the state.
  • Displayed arrows can’t be rebuilt from the readings: a least-squares slope of 4 readings in 15–20 min matched the Dexcom G5’s displayed arrow in 90.7 % of cases, two readings matched the Libre 1’s in 77.9 %; “the displayed trend arrows cannot be reliably reconstructed based on the stored CGM readings” (Eichenlaub, Pleus, Freckmann 2024, PMC11307235).
  • The FDA’s iCGM rule judges the rate only by its worst errors: ≤ 1 % of readings may say > +1 when the true rate is < −2 mg/dL/min, and the reverse (21 CFR 862.1355 (b)(1)(i)(J)/(K)).
  • Physiology: rates over 2 mg/dL/min about 10 % of the time around meals (Dunn 2004, Diabetes Care 27:2161); sustained rates of ±2 / 3 / 4 mg/dL/min as 90 / 95 / 99 % bounds (Kovatchev 2005, DTT 7:849, from an excerpt).
  • Savitzky–Golay: no CGM paper on SG rate estimation found; xDrip+ and xDrip4iOS use SG to smooth the drawn values.

DIY apps (code read at each repo’s HEAD, 2026-11-14):

App Trend Source
xDrip+ two-point slope of the newest two readings (1 min apart with Libre 1-min data); bands ±1 / 2 / 3.5; SG smoothing for the graph only; Libre 2 values: a linearly decaying weighted average over 25 min BgReading.java L305-311, L1734-1749, Dex_Constants.java L86-155, LibreReceiver.java L138-149
xDrip4iOS two-point slope against the nearest reading 4–21 min older (“very close readings … become an extreme per-minute slope”); optional SG value smoothing (two passes for 1-min data) BgReading+CoreDataClass.swift L280-311, ConstantsBgSmoothing.swift
Nightscout delta: mean of the newest 5-min bucket − mean of the one before; the arrow is the uploader’s direction (no bands of its own) bgnow.js L142-172, direction.js L48
oref0, Trio delta = mean of (now − then) / minutes × 5 for readings 2.5–7.5 min old; short_avgdelta 2.5–17.5; long_avgdelta 17.5–42.5. Fed 1-min readings unbucketed it degenerates: its running now_date keeps every next reading < 2.5 min away, so all are averaged into “now” and the deltas are 0 (found by the prototype) glucose-get-last.js L25-84, Trio DetermineBasal+Helpers.swift L52-85
AndroidAPS the same windows (DeltaCalculator) over readings first interpolated onto a 5-min grid ending at the newest one; AutoISF fits a parabola over up to 47.5 min; optional smoothing plugins incl. an adaptive unscented Kalman filter with state [G, Ġ] DeltaCalculator.kt L46-113, AutosensDataStoreObject.kt L331-433, UnscentedKalmanFilterPlugin.kt L47-60
Loop glucose momentum: ordinary least squares over 15 min, > 2 readings, continuous, one source, no calibrations, capped at 4 mg/dL/min; the arrow is the CGM’s own GlucoseMath.swift L65-118
Juggluco the sensor’s rate; the drawn arrow is continuous, atan(0.75 × rate), near-zero snapped flat; sensors without a rate: a weighted regression of the newest 3 points curve.cpp L1701-1740, interpretgs3.cpp L376-393
GlucoDataHandler the source’s rate; optionally its own: per-reading slopes over 25 min weighted 2 / 1.5 / 1 / 0.5 by age; rotated arrow clamp(rate, ±2) / 2 × 90°, ±14° dead zone, 5° steps ReceiveData.kt L423-443, GlucoDataUtils.kt L176-182

Nobody publishes a comparison of these on real data. The DIY apps that compute a trend use a two-point delta or an average of deltas; only research uses Kalman filtering.

  • “the direction of the trend arrow does not match the actual trend”; a maintainer: it is “calculated using the current and previous reading”; “I’ll see my bg is stable until I investigate the graph” (one equal reading during a drop) (xDrip+ discussion #3861, 2025-26).
  • Libre 3 in xDrip+: “a minute time delta would lead to a non significant bg delta”, and smoothing on top “leads to a big delay” (#2598, 2023). Libre 3 1-min values “pretty raw and jumpy at times” (#4102, from a search excerpt).
  • The owner (#193, #204): the sensor’s arrow said falling minutes into the rebound and flat during a clear rise.
  • Reddit threads could not be fetched (blocked) and searches returned none: not verified.

Both failure modes show up: too jumpy (two-point slopes) and too late (the sensor’s own smoothing).

All causal and run on every reading, as the apps would (methods.py): two-point delta (newest vs the reading 1 min / 5 / 10 / 15 min before: xDrip+ and Nightscout); AndroidAPS delta and shortAvgDelta (5-min grid, as above); least squares over 4–15 min (Sukrara’s #193 Trend is 5 min, Loop 15 min); recency-weighted least squares exp(−age / τ), τ 2–5 min; local quadratic evaluated at the newest reading (the Savitzky–Golay end-point derivative), 7–30 min, plain and weighted; Theil–Sen (median of pairwise slopes; robust to one outlier) over 5 / 10 min; Holt double exponential smoothing for uneven steps (α 0.2–0.7, β 0.1–0.3 per minute); a constant-velocity Kalman filter (white-noise acceleration q 0.001–0.05 (mg/dL/min²)²/min, reading noise 1 mg/dL², restart after 10 min); and Abbott’s rate as recorded. Each window method restarts after a gap and needs a minimum of readings; evaluate.py --sweep runs all 50 parameter sets.

Data: 6,501 Libre 3 readings, one a minute, 2026-11-10 to 2026-11-14 (4.6 days), with Abbott’s rate for each; 4 holes over 5 min; no LO / HI; read locally from the Mac’s readings.csv, nothing derived from them is committed except the aggregates below. Reading-to-reading noise is about 0.6 mg/dL, but steps of 5 mg/dL or more from one minute to the next happen about 70 times a day (7 or more: about 25).

“Truth”: a hindsight slope the apps can’t have: at each reading, the slope of a parabola fitted with tricube weights to the readings ±7 min around it (a centred Savitzky–Golay derivative). It is the slope of what the person sees, not of blood glucose. Turning points: 203 times it crosses 0, 145 times ±1 and 64 times ±2 (each held ≥ 5 min).

Metrics: delay (best fit): the shift that best lines the estimate up with the truth, and the gain there (< 1: the arrow understates the move); delay to cross 0 / ±1 / ±2: minutes after the truth until the estimate does, median (p90); missed: crossings it didn’t make within 20 min (of 412); band changes per hour of the Nightscout bands (±1 / 2 / 3), raw and with hysteresis 0.2; angle change: the median change of the continuous arrow (§6) from one reading to the next. The truth itself changes band 9.3 times an hour (7.9 with hysteresis), so a perfect, delayless method would too.

1-min readings (evaluate.py READINGS):

Method Delay Gain Cross 0 Cross ±1 (p90) Cross ±2 Missed Band changes/h … hysteresis 0.2 Angle change
Abbott’s rate (today) 3.5 min 0.35 5 7 (17) 15 214 1.5 1.0 1.2°
delta, newest two (xDrip+) 0.5 1.30 −2 −1 (1) 0 0 33.0 17.6 29.5°
delta 5 min (Nightscout; AAPS delta) 2.5 1.16 2 3 (4) 3 0 14.0 11.7 6.0°
AAPS shortAvgDelta 3.5 0.82 4 4 (6) 4 5 7.6 5.9 4.1°
least squares 5 min (#193 quiets) 2.0 1.22 1 2 (3) 2 0 16.9 15.0 8.9°
least squares 15 min (Loop) 7.5 0.78 7 7 (9) 7 9 5.4 4.2 2.8°
quadratic 15 min (S-G end point) 3.5 1.56 3 3 (4) 3 0 17.5 15.5 7.5°
Theil–Sen 10 min 4.5 1.02 4 5 (5) 5 0 9.0 7.5 3.8°
Holt α 0.5 β 0.2 3.5 0.91 4 4 (5) 4 1 8.3 6.6 4.2°
Kalman q 0.01 4.0 0.95 4 4 (5) 4 0 8.5 7.0 4.2°
weighted LS τ 2 (Quick) 2.5 1.08 3 3 (4) 3 0 12.7 10.2 6.1°
weighted LS τ 3 (Balanced) 3.5 0.90 4 4 (5) 4 0 8.1 6.5 4.1°
weighted LS τ 5 (Smooth) 5.5 0.65 6 6 (10) 6 80 3.8 2.8 2.2°

What it says:

  • Abbott’s rate is not just late, it is flat. It moves at about a third of the readings’ slope, so it reaches ±1 late or never (214 of 412 crossings missed within 20 min) and almost never says ↑ / ↓. That is the owner’s experience in both episodes. Its low band-change count is the same flatness, not quality.
  • Every honest method trades delay for band changes along one line. At about 6–7 band changes an hour (close to the truth’s own 7.9) the recency-weighted slope, Holt, the Kalman filter and AAPS’ shortAvgDelta all reach ±1 in 4 min; plain least squares needs a 12-min window for the same calm and is then 1 min slower (sweep: 5 min at 6.1 / h). Weighting the newest readings more is what buys the minute.
  • The two-point delta of the newest readings is unusable at 1 min (33 band changes an hour, the arrow turning 30° a minute), and the 5-min delta and the 5-min least squares change band twice as often as τ 3 for 1–2 min less delay.
  • The Savitzky–Golay end point has the least lag for a smooth curve in theory and the worst noise in practice: more band changes than the 5-min slope at the same delay, and it overshoots (gain 1.56).
  • Theil–Sen scores like plain least squares of the same window: a robust slope buys nothing here, because the Libre 3’s jumps are steps that stay, not single readings off the line.

Robustness (all in evaluate.py):

Test τ 3 (Balanced) Others
Truth ±5 / ±10 min instead of ±7 ±1 in 4 (6) / 4 (5) min same ranking: delta 5 min 3 / 2 min with 2× the band changes, Loop 7 / 7, Abbott 6 (15) / 8 (17), gain 0.29 / 0.45
Extra reading noise sd 2 mg/dL (rounded) ±1 in 4 (5); 7.3 band changes/h with hysteresis delta 5 min 17.9 / h, least squares 5 min 19.4, newest-two delta 44.7, AAPS short 7.9, Holt / Kalman 7.5
40 random 3–20 min holes a rate on 99 % of judged readings; after a hole: at once if ≤ 12.5 min (bridged), else after 3 min of readings (max 6) Loop-style 15-min continuity: median 8, max 37 min without an arrow
5-min sensor (every 5th reading) ±1 in 5 (8) min; 4.2 band changes/h; same parameters delta 5 min 5 (7) at 5.2 / h; AAPS short 6 (9), 34 missed; Loop 7 (9); Holt 6 (8); Kalman q 0.01 7 (10)
Hysteresis 0 / 0.1 / 0.2 / 0.3 8.1 / 7.3 / 6.5 / 5.9 band changes/h; entering ±1 on average 4.2 / 4.5 / 4.8 / 5.1 min after the truth —

The two episodes (--episode; in words only):

  • The #193 evening, 20:08–20:45 (after holes at 20:02 and 20:19): the readings’ hindsight slope turned upward around 20:28, passed +1 at 20:29 and +2 at 20:33, peaking near 3.4 mg/dL/min. Abbott’s rate turned positive only at 20:38, reached +1 at 20:44 (15 min late) and never went above 1.0: the arrow never showed more than ↗. τ 3 showed a rise as soon as 3 min of readings followed the 20:19 hole, passed +1 at 20:33 and +2 at 20:37, peaking near the truth’s 3.4.
  • The rise of 2026-11-14 afternoon (14:30–15:10): the hindsight slope was positive from about 14:35 and above +1 from 14:45, peaking near 1.8. Abbott’s rate turned positive at 14:46 and peaked at 0.8: → for the whole rise. τ 3 was positive at 14:39, passed +1 at 14:48 and peaked at 1.8.

The method. At every reading, from the readings this device has:

  1. Walk back from the newest reading while readings are less than 15 min (5τ) older than it, stopping at a LO / HI or at a step longer than 12.5 min between two readings (2.5 readings of a 5-min sensor; a 1-min sensor’s shorter holes are bridged: the slope across them is still a slope).
  2. Need at least two readings spanning at least 3 min, else no trend from the readings.
  3. Weight each reading w = exp(−(t_newest − t) / τ), τ = 180 s; time in seconds from the newest reading; sums taken oldest first; rate = Σw·(dt − t̄)·(v − v̄) × 60 / Σw·(dt − t̄)² mg/dL/min (t̄, v̄ the weighted means).

Why this one over its equals: it is a pure function of the readings (Holt and Kalman carry state: a restart, a backfilled or late reading, a reordered sync or a follower joining later would each get a different arrow; their “after a gap” is a reset with its own warm-up), it is defined in time, not in readings (AAPS’ deltas need a 5-min grid and interpolation; the same τ is right for 1-min and 5-min sources, §4), it is exact on a straight line and lags a steady acceleration by about 4.5 min instead of plain 15-min least squares’ 7, and it is the same arithmetic as the #193 Trend slope with weights added, so both platforms already sum it the same way.

Presets. The line from quick to smooth is continuous (§4) and the right point is personal (someone who doses from the arrow wants it quick; someone who glances wants it calm). One setting, four choices: Quick (τ 2 min: ±1 in 3 min, 10.2 band changes/h), Balanced (τ 3, the default: 4 min, 6.5 / h), Smooth (τ 5: 6 min, 2.8 / h, starts missing short moves), From the sensor (the sensor’s own rate, the owner’s existing decision). These replace the plan’s least-squares windows “Quick 4 / Balanced 5 / Smooth 10 min” (sweep, hysteresis 0.2): Quick is as calm as an 8-min window (10.2 / h) and reaches ±1 a minute sooner; Balanced is calmer than the 10-min window (6.5 vs 8.0 / h) and turns a minute sooner (crossing 0: 4 vs 5 min); a 4- or 5-min window changes band 15–16 times an hour.

Text bands (text arrows, spoken words, Nightscout direction, the xDrip+ broadcast’s own rule aside): Nightscout’s ±1 / 2 / 3 with hysteresis 0.2 mg/dL/min: a band is left only once the rate is 0.2 past its edge (→ to ↗ at 1.2, back to → below 0.8). The band remembered is per device and starts fresh with no previous band.

Fallbacks: the readings’ trend → the newest reading’s own rate from the sensor (when it has one) → no trend (the “no trend” ring). LO / HI as the newest reading: no trend from the readings; the readings before a LO / HI never count. The freshness rules of the card stay as they are (a stale value already hides or marks its arrow).

Cost: at most 15 readings (1-min) or 3 (5-min): 15 exp and about 60 multiply-adds per reading, 13 µs in CPython, well under a microsecond compiled; memory: the last 15 min of readings (the alert Trend keeps 5 min today).

Shared test vectors: vectors.py writes the proposal, synthetic readings only: flat, straight rises and falls at 1 and 5 min, uneven timing, alternating noise, an acceleration (shows the 4.5-min lag), a turn, too few readings, exactly 3 min, a 5-min sensor’s second reading, an 8-min hole bridged, a 13-min gap, a newest LO, readings after a HI, an old rise outside the window, duplicates and disorder, a compression-like dip; then a sequence of rates with the bands they must show under hysteresis, and the drawn angles. Rates compare within 1e-9 (exp may differ in the last bit between platforms).

  • Mapping: linear, deg = clamp(rate / 3, −1, 1) × 90° (the PoC; y-down: up is negative), doubled from |rate| ≥ 3 with the same hysteresis as the bands. With the bands at ±1 / 2 / 3, the drawn arrow is always within 30° of the text arrow beside it: → covers −30…30°, ↗ 30…60° (its symbol is 45°), ↑ 60…90°. Juggluco’s atan(0.75 × rate) never reaches vertical (66° at 3, so “rising fast” at 2 is drawn 56°, 34° away from ↑); GlucoDataHandler’s ±2 → vertical with a ±14° dead zone and 5° steps is a quantised version of the same idea.
  • No smoothing of the angle. τ 3 already moves the arrow a median 4° per 1-min reading (p90 15°, p99 32°; the hindsight slope itself 4.4°): any further filter is extra delay on the one thing the change is for. Animate each change with a short spring without overshoot (Material motion / SwiftUI spring, ~300 ms); reduced motion: jump. A dead zone around flat is not needed (a 3° tilt at 0.1 mg/dL/min is the truth, and small).
  • Places that can’t rotate (system icons, widgets) keep pre-rendered steps; the PoC’s 5° steps are fine-grained enough.
  • The arrow will say ↑ / ↓ and ⇈ / ⇊ far more often. The owner’s readings move at ≥ 2 mg/dL/min about 17 times a day even by the ±10-min hindsight slope (median 6 min each); Abbott’s rate says so about twice a day, τ 3 about 25 times (14 if held 5 readings), τ 5 about 11. ≥ 3 (doubled arrow): Abbott ~0.6 a day, τ 3 ~12. That is the readings being honest, but it is a visible change; Smooth exists for those who prefer calm.
  • The rate alerts can’t simply switch to this trend. falling_fast / rising_fast (off by default, 2.0 mg/dL/min) would go from about 2 episodes a day to about 25 on these readings. Before they read the one trend they need their own rule (for example the rate held for 10 min: ~3.5 a day, close to the hindsight slope’s 3.5) and new timeline scenarios, with the owner’s OK (CLAUDE.md: never change a scenario’s expectations without it).
  • The #193 quiets keep their own rule. Their 5-min least squares is the fastest smooth-enough slope (±1 in 2 min) and their extra guards (no step against the direction over 1 mg/dL, still moving, at least 0.5 mg/dL/min) answer a different question: “is it clearly still moving?” A weighted slope would notice a stall 1–2 min later, the wrong direction for a rule that holds an alert. Nothing here is clearly better for them. Their noise guard is conservative with the Libre 3’s 5–10 mg/dL steps (the step voids the trend and the alert fires as without the quiet): the safe side.
  • Compression lows: the arrow follows the readings’ fall and rebound like the value does, within 3–4 min; docs/33 stands (nothing at the alarm says “pressure”).
  • Followers compute from the readings they have: a 5-min Nightscout follower gets the 5-min numbers of §4 with the same parameters; the wire keeps the sensor’s rate; LibreView / Tidepool / CSV keep the sensor’s rate (owner’s decision).
  1. Presets: all four, one setting (Quick / Balanced / Smooth / From the sensor).
  2. The doubled arrow: pick the clearly better rule from the data (below: held 2 min).
  3. The rate alerts: move them onto the trend now, with their own rule from the data, all the way (below: held 5 min).
  4. Hysteresis 0.2 for the text bands: yes; the drawn arrow stays continuous.

Built in #204 (both apps, one round). The numbers below are investigations/204-trend/holds.py run over the owner’s 1-min readings (4.7 days by then; aggregates only); “truth” is the ±10-min hindsight slope of §4.

Brought in Why
THE trend: TrendRate (Android :core:model, Mac SukraraCore) = §5’s weighted least squares, τ from the setting, fallback readings → the sensor’s rate → none; each reading’s trend at its own time from the readings at or before it, worked out at one place per platform (Android TrendReadings, the readings repository every reader goes through; Mac AppModel); Reading.sensorRate is the stored / wire / device-record rate, Reading.trend the one every arrow, word, output and alert reads the owner’s “one source of truth”; §4: ±1 in 4 min instead of 7 (17), no missed turns
The setting Trend arrow (glucose.trend: quick / balanced (default) / smooth / sensor, mirrored as graph.trend), on Android Settings › Ranges › Trend arrow (docs/35 moves it to Glucose), on the Mac Settings › General; a live preview (glyph surface trend-preview) §5 presets; the owner’s choice of four
Text bands ±1 / 2 / 3 with hysteresis 0.2, the band remembered while readings follow within 12.5 min with a trend §4: −20 % band changes for +0.7 min
The doubled arrow held 2 min: band ±3 is entered only when the rate was ≥ 3 at every reading back to the first at or before 2 min earlier (3 readings of a 1-min sensor, 2 of a 5-min one), left with the 0.2. Balanced, measured: plain ≥ 3 with hysteresis 11.4 doubled arrows a day (0.4 a day with no truth ≥ 2.5 near it, 59 doubled min a day), held 1 min 9.7 (0.4), held 2 min 7.5 (0.2, 35 min a day), held 3 min 6.7 (0.2); each misses 1 of the 20 truth moves ≥ 3 held 3 min; median delay to them 4 → 6 min (p90 5 → 7; 3 min: 6, p90 8). Quick: 22.5 → 15.7 a day; Smooth: 2.4 → 2.1 the knee: 2 min (3 readings of a 1-min sensor) takes a third of the doubled arrows and half the false ones away for 2 min; a third minute buys 0.8 a day for another minute at p90. A plain 3.5 (no hold) keeps the short false ones
The drawn arrow continuous (deg = −clamp(rate / 3, −1, 1) × 90), doubled with the band, a ~300 ms spring without overshoot on the Android home card and the settings previews (none with reduced motion); system icons draw the band §6
The fast alerts held 5 min (SPEC §2 Fast alerts read the trend): held = over the readings back to the first at or before 5 min earlier, all with a trend, the rate nearest zero (0 on mixed signs, none when they don’t reach back 5 min); From the sensor: the sensor’s rate at once, as before. At 2.0 (default) on Balanced: no hold 17.2 alert episodes a day (30-min episodes; ~25 counting every swing), held 2 min 15.7, 3 min 13.5, 5 min 9.4 (0.4 a day without a move ≥ 1.5 nearby, median 9 min after the truth’s crossing, p90 10, missing 1 of 39 truth moves ≥ 2 held 5 min), 10 min 2.6 (missing 28); a 5-min mean of rates 13.6, a 10-min mean 8.0 (missing 4), a 15-min delta + now 9.7 (missing 3) (first pass, not in holds.py). Truth: 8.4 such moves a day. Abbott’s rate (before): 1.9 a day, 11 min (p90 19), missing 32 of 39; held 5 min it would be 1.3 a day, 16 min (p90 24). At 3.0: held 5 min 3.2 a day (truth 2.1), missing 1 of 10. Quick: 10.9 a day; Smooth: 6.4 (missing 16) as often as glucose really moves that fast for 5 min, almost never without such a move, and far sooner and more complete than the sensor’s rate; the simplest rule at that point. Holding the sensor’s rate too would add 5 min to a rate that already lags 7
Rate-alert scenarios: rate.txt (a real fall with a lagging sensor, flat readings with the sensor at −3, the same from the sensor, a 5-min sensor, a gap, switching the setting mid-fall, a mirrored follower, the owner’s evening rising fast from the readings at 20:45 and never from the sensor); the scenarios about LO / carbs / mute / no-data interplay with the fast alerts pinned to trend=sensor (their expectations unchanged: they test the interplay, not the trend); the format gains glucose … moving <per min> and trend= the owner’s approval to change the rate scenarios; CLAUDE.md alert-scenario rule
The xDrip+ broadcast takes the hub’s trend for the sensor’s live reading (keeping xDrip+’s 3.5 doubling); the local Nightscout / xDrip+ endpoints and the Nightscout upload send the trend’s direction and delta, none without a trend (their three two-point slopes and the “0 for unknown” are gone) one source of truth; a 0 delta reads as flat on a watch
contracts/trend/ (rate.json per reading, bands.json, clear.json = the #193 quiets’ cases, moved) written by tools/trend/cases.py (grown from vectors.py and tools/alerts/trend_cases.py), run by both apps; TrendEverywhereTest (Android) / TrendEverywhereTests (Mac): the owner’s rise through the one place, the same arrow on every surface; the repo check trend pinned like the alert scenarios; the regexes catch reads and writes outside the module, the cross-surface tests the rest
docs/FEATURES.md Trend arrow with its states and exits; QA Q26 complete flows
Left out Why
Kalman / Holt filters the same numbers as the weighted slope (§4) with state: restarts, backfill and followers would disagree
Savitzky–Golay end-point and other local-quadratic slopes the noisiest at every delay; overshoot
Two-point deltas (xDrip+, Nightscout) and AAPS’ deltas 2× the band changes, or a 5-min grid; oref0 unbucketed returns 0 on 1-min data
Theil–Sen no gain: the Libre 3’s jumps are steps, not outliers
A per-person tuned filter (Facchinetti) needs per-sensor noise estimation; the presets cover the preference
Smoothing or a dead zone for the drawn angle extra delay for a 4° median change per reading
An option for the doubled arrow or the alerts’ hold the data shows one clearly better point each; nothing says people want different behaviours
Changing the #193 quiets §7: their rule is faster where it matters and errs to alerting; moved beside the trend unchanged (TrendRate.clear)
Using this trend for LibreView / Tidepool / CSV or on the wire the owner’s decision: those keep the sensor’s rate (device records); followers work it out from their readings
A spring on the Mac’s menu bar, dropdown and widgets they are drawn as images; the arrow jumps there (the settings preview springs)
Real readings, per-reading outputs or plots in the repo, beyond the two episodes the owner allowed in the cases (#193 evening, the 2026-11-14 rise) health data stays on the owner’s machine